H-3301/H-3311/H-3342 Shaft Encoder
The WaterLOG® 3300 series digital shaft encoders measure water level by monitoring the position of a float and pulley. They are easy to use and work with any SDI-12 data recorder.
DAA has Authorized Distributors all over the globe. View a map and find your closest contact for DAA H-3301/H-3311/H-3342 Shaft Encoder products and accessories.
Have a question about your H-3301/H-3311/H-3342 Shaft Encoder? Check out the support page with resources compiled just for you.
Want to return a product for service? Here you'll find authorized service centers and instructions on how to send in your H-3301/H-3311/H-3342 Shaft Encoder.
General
• Battery powered operation with minimal current drain
• Low system cost
• Multiple sensors on a simple three-wire cable (SDI-12)
• Magnetic sensing minimizes static sensitivity
• Built-in LCD Display with "Push to Read" Button (3342 and 331 only)
• Threaded shaft compatible with older mechanical shaft encoders
H-3342
• Absolute, non-contact, optical encoder preserves correct position even if the power is lost.
• Resolution = 65,536 counts/rev
• Accuracy = 0.00024 ft. (1 ft. diameter pulley)
• Max turns = ±32,768
• Zero backlash
• 4-20mA output
• Precision dual bearing design with special low temperature lubricant
H-3301/3311
• 200 counts per revolution for the SDI-12 output
• 100 counts per revolution for the quadrature output
Applications
• Scales the encoder position into user units of Feet, Meters, etc.
| Data Input | |||
|
General
|
Inputs | Shaft Position | |
|
Outputs |
SDI-12 & 2-Wire Quadrature |
||
| Resolution (SDI) | 200 counts/rev | ||
| Resolution (Quadrature) | 100 counts/rev | ||
| Max Shaft Rotation | 15 rev/sec | ||
| |
|||
| Data Output | |||
| SDI-12 Output
|
Baud Rate | 1200 | |
|
Protocol |
SDI-12 7-bit even parity, 1 stop bit |
||
|
Output Voltage Level: |
Minimum high level: 3.5 volts |
||
|
Maximum low level: 0.8 volts |
|||
|
Maximum cable length: 250 ft. |
|||
|
Quadrature Output
|
Type | Open Drain (optional factory installed pull up resistor to 5.0 V) | |
|
Transient Protection |
5.0 V Transguard |
||
| |
|||
| Power Requirements | |||
|
|
Voltage Input | 9.6 to 16.0 Volts DC | |
|
Current
|
Less than 2.0mA |
||
| 800μA typical | |||
| |
|||
| Mechanical | |||
|
Physical
|
Bearing |
Double bearing with external seal |
|
|
Starting Torque
|
0.15 oz in typical |
||
|
0.50 oz in max over temperature |
|||
| Shaft |
5/16 in. compatible with Stevens and Fisher and Porter pulley and accessories |
||
|
Shaft Length |
2 in |
||
|
Shaft Thread |
0.75 in |
||
| Material |
Anodized Aluminum |
||
|
Size
|
5.125 in wide (main unit) |
||
|
7.0 in wide (base plate) |
|||
|
4.25 in high |
|||
|
4.0 in deep (not including shaft or connectors) |
|||
|
Connector
|
H-3301 | Amphenol MS3102R14S-6P (6-Pin male) | |
| Cable | Amphenol MS3106A14S-6S (6-Pin female) | ||
| (One 6 foot cable provided with the H-3301 | |||
| Environmental | |||
|
|
Operating Temperature | -40° C to 60° C | |
|
Storage Temperature |
-50° C to 70° C | ||
|
Humidity |
0 to 100% |
||
| Miscellaneous | |||
| Warranty |
The WaterLOG® H-3301 is warranted against defects in materials and workmanship for two years from date of shipment. |
||
|
Specifications subject to change without prior notice due to ongoing commitment to product testing and improvement. LR Jan 10 2008 |
|||
|
|
H-3301-CABLE Communication CableH-3301-CABLE Communication Cable for H-3301 & H-3311 Shaft Encoders |
|